McMillan Model 50D Thermal Mass FLO-METERS are capable of measuring virtually any clean,dry gas as low as 0-20 sccm or as high as 0-500 Lpm! Repeatable results are achieved using a patented thermal mass flow sensor design. This proven design minimizes zero drift while maintaining fast response and linear outputs.
Because of the compact size and economical cost of these products, the Model 50D Series FLO-METERS are suitable for a wide variety of industrial, commercial, laboratory and O.E.M. applications.
An integrated display reports flow in either mL/minute or L/minute. An analog output is also provided for remote data logging.
A cable assembly or power adapter is required for operation.
Model 50D Features & Options
ACCURACY/LINEARITY
All models have a standard accuracy specification of ±1.5% F.S. accuracy (including linearity). NIST traceable calibration certificates are optional on all models.
CALIBRATION GASES
Units may be calibrated for virtually any clean, dry gas. Several standard gas selections are available as indicated in the Product Configurator. Contact factory for calibration information on non-standard gases.
DISPLAYS
A 3.5 digit LCD flow rate display is integrated, which displays current flow rate in either mL/minute or L/minute. The display angle may be adjusted. If totalizing or other functions are required, consider McMillan’s line of external remote displays.
ELECTRICAL CONNECTIONS
All units have a 36″ (92 cm) output cable, terminated with either a 9-pin “D” connector, 15-pin “D” connector, or 6-pin PS/2 style connector. An optional mating cable assembly, terminated with pigtail leads, is recommended to facilitate wiring.
FLOW RANGES
Flow ranges from 0-20 sccm up to 0-500 L/minute are available. Consult the factory for custom requirements.
FLUID CONNECTIONS
All units have compression-type tube fittings as standard. Consult the Product Configurator for available materials and sizes.
POWER
Units may be ordered to operate with either 12 VDC or 24 VDC power. Various power adapters are available for use with 12 VDC versions.
SIGNAL OUTPUTS
All integrated output signals are 0-5 VDC. An external module may be ordered to provide 4-20 mA output in addition to the integrated 0-5 VDC output.
WETTED MATERIALS
All units feature metal construction. See specifications for detailed materials in gas path. 50S and 50SD units feature stainless steel construction; 50K and 50D units are constructed from aluminum.
Model 50K
McMillan Model 50K Thermal Mass FLO-SENSORS are capable of measuring virtually any clean, dry gas as low as 0-20 sccm or as high as 0-500 Lpm! Repeatable results are achieved using a patented thermal mass flow sensor design. This proven design minimizes zero drift while maintaining fast response and linear outputs.
Because of the compact size and economical cost of these products, the Model 50K Series FLO-SENSORS are suitable for a wide variety of industrial, commercial, laboratory and O.E.M. applications.
An integrated analog output reports flow rate in VDC proportional to flow rate.
A cable assembly or power adapter is required for operation.
50K Features & Options
ACCURACY/LINEARITY
All models have a standard accuracy specification of ±1.5% F.S. accuracy (including linearity). NIST traceable calibration certificates are optional on all models.
CALIBRATION GASES
Units may be calibrated for virtually any clean, dry gas. Several standard gas selections are available as indicated in the Product Configurator. Contact factory for calibration information on non-standard gases.
DISPLAYS
For units with integrated displays, consider the Model 50D or 50SD. Units without integrated displays may be used with McMillan’s line of external remote displays.
ELECTRICAL CONNECTIONS
All units have a 36″ (92 cm) output cable, terminated with either a 9-pin “D” connector, 15-pin “D” connector, or 6-pin PS/2 style connector. An optional mating cable assembly, terminated with pigtail leads, is recommended to facilitate wiring.
FLOW RANGES
Flow ranges from 0-20 sccm up to 0-500 L/minute are available. Consult the factory for custom requirements.
FLUID CONNECTIONS
All units have compression-type tube fittings as standard. Consult the Product Configurator for available materials and sizes.
POWER
Units may be ordered to operate with either 12 VDC or 24 VDC power. Various power adapters are available for use with 12 VDC versions.
SIGNAL OUTPUTS
All integrated output signals are 0-5 VDC. An external module may be ordered to provide 4-20 mA output in addition to the integrated 0-5 VDC output.
WETTED MATERIALS
All units feature metal construction. See specifications for detailed materials in gas path. 50S and 50SD units feature stainless steel construction; 50K and 50D units are constructed from aluminum.
Model 50S
McMillan Model 50S Thermal Mass FLO-SENSORS are capable of measuring virtually any clean, dry gas as low as 0-20 sccm or as high as 0-100 Lpm! Repeatable results are achieved using a patented thermal mass flow sensor design. This proven design minimizes zero drift while maintaining fast response and linear outputs.
Because of the compact size and economical cost of these products, the Model 50S Series FLO-SENSORS are suitable for a wide variety of industrial, commercial, laboratory and O.E.M. applications.
An integrated analog output reports flow rate in VDC proportional to flow rate.
A cable assembly or power adapter is required for operation.
50S Features & Options
ACCURACY/LINEARITY
All models have a standard accuracy specification of ±1.5% F.S. accuracy (including linearity). NIST traceable calibration certificates are optional on all models.
CALIBRATION GASES
Units may be calibrated for virtually any clean, dry gas. Several standard gas selections are available as indicated in the Product Configurator. Contact factory for calibration information on non-standard gases.
DISPLAYS
For units with integrated displays, consider the Model 50D or 50SD. Units without integrated displays may be used with McMillan’s line of external remote displays.
ELECTRICAL CONNECTIONS
All units have a 36″ (92 cm) output cable, terminated with either a 9-pin “D” connector, 15-pin “D” connector, or 6-pin PS/2 style connector. An optional mating cable assembly, terminated with pigtail leads, is recommended to facilitate wiring.
FLOW RANGES
Flow ranges from 0-20 sccm up to 0-100 L/minute are available. Consult the factory for custom requirements.
FLUID CONNECTIONS
All units have compression-type tube fittings as standard. Consult the Product Configurator for available materials and sizes.
POWER
Units may be ordered to operate with either 12 VDC or 24 VDC power. Various power adapters are available for use with 12 VDC versions.
SIGNAL OUTPUTS
All integrated output signals are 0-5 VDC. An external module may be ordered to provide 4-20 mA output in addition to the integrated 0-5 VDC output.
WETTED MATERIALS
All units feature metal construction. See specifications for detailed materials in gas path. 50S and 50SD units feature stainless steel construction; 50K and 50D units are constructed from aluminum.
Model 50SD
McMillan Model 50SD Thermal Mass FLO-METERS are capable of measuring virtually any clean, dry gas as low as 0-20 sccm or as high as 0-100 Lpm! Repeatable results are achieved using a patented thermal mass flow sensor design. This proven design minimizes zero drift while maintaining fast response and linear outputs.
Because of the compact size and economical cost of these products, the Model 50SD Series FLO-METERS are suitable for a wide variety of industrial, commercial, laboratory and O.E.M. applications.
Stainless steel construction and standard metal fittings provide reliable service for most applications.
An integrated display, showing flow rate in mL/minute or L/minute, is combined with an analog output for data recording.
A cable assembly or power adapter is required for operation.
50SD Features & Options
ACCURACY/LINEARITY
All models have a standard accuracy specification of ±1.5% F.S. accuracy (including linearity). NIST traceable calibration certificates are optional on all models.
CALIBRATION GASES
Units may be calibrated for virtually any clean, dry gas. Several standard gas selections are available as indicated in the Product Configurator. Contact factory for calibration information on non-standard gases.
DISPLAYS
A 3.5 digit LCD flow rate display is integrated, which displays current flow rate in either mL/minute or L/minute. The display angle may be adjusted. If totalizing or other functions are required, consider McMillan’s line of external remote displays.
ELECTRICAL CONNECTIONS
All units have a 36″ (92 cm) output cable, terminated with either a 9-pin “D” connector, 15-pin “D” connector, or 6-pin PS/2 style connector. An optional mating cable assembly, terminated with pigtail leads, is recommended to facilitate wiring.
FLOW RANGES
Flow ranges from 0-20 sccm up to 0-100 L/minute are available. Consult the factory for custom requirements.
FLUID CONNECTIONS
All units have compression-type tube fittings as standard. Consult the Product Configurator for available materials and sizes.
POWER
Units may be ordered to operate with either 12 VDC or 24 VDC power. Various power adapters are available for use with 12 VDC versions.
SIGNAL OUTPUTS
All integrated output signals are 0-5 VDC. An external module may be ordered to provide 4-20 mA output in addition to the integrated 0-5 VDC output.
WETTED MATERIALS
All units feature metal construction. See specifications for detailed materials in gas path. 50S and 50SD units feature stainless steel construction; 50K and 50D units are constructed from aluminum.
Fitting Chart
Specifications
- Accuracy (including linearity, best fit straight line):
- Model 50D: ±1.5% Full Scale (10-100% rated flow)
Model 50K: ±1.5% Full Scale (10-100% rated flow)Â
Model 50S: ±1.5% Full Scale (10-100% rated flow)Â
Model 50SD: ±1.5% Full Scale (10-100% rated flow) - Repeatability:
- Model 50D: ±0.25% Full Scale (10-100% rated flow),
Model 50K: ±0.25% Full Scale (10-100% rated flow) ,
Model 50S: ±0.25% Full Scale (10-100% rated flow) ,
Model 50SD: ±0.25% Full Scale (10-100% rated flow), - Pressure Rating:
- Model 50D: 150 psig (10.3 bar)
Model 50K: 150 psig (10.3 bar)Â
Model 50S: 500 psig (34.5 bar)Â
Model 50SD: 500 psig (34.5 bar) - Pressure Sensitivity:
- Model 50D: ±0.02% Full Scale per psi (per 69 mbar)
Model 50K: ±0.02% Full Scale per psi (per 69 mbar)Â
Model 50S: ±0.02% Full Scale per psi (per 69 mbar)Â
Model 50SD: ±0.02% Full Scale per psi (per 69 mbar) - Temperature Rating (Environment):
- Model 50D: Operating: 5 to 55ºC
Storage: 0 to 70ºC
Model 50K: Operating: 5 to 55ºC
Storage: 0 to 70ºC
Model 50S: Operating: 5 to 55ºC
Storage: 0 to 70ºC
Model 50SD: Operating: 5 to 55ºC
Storage: 0 to 70ºC - Temperature Sensitivity:
- Model 50D: ±0.15% F.S. or less per ºC Model 50K: ±0.15% F.S. or less per ºC Model 50S: ±0.15% F.S. or less per ºC Model 50SD: ±0.15% F.S. or less per ºC
- Body Leak Integrity (not including fittings):
- Model 50D: 1 x 10-7 sccs of He Model 50K: 1 x 10-7 sccs of He Model 50S: 1 x 10-7 sccs of He Model 50SD: 1 x 10-7 sccs of He
- Wetted Materials:
- Model 50K: Aluminum
304 Stainless Steel
316 Stainless Steel
Model 50S: 303 Stainless Steel
304 Stainless Steel
316 Stainless Steel
Epoxy
Model 50SD: 303 Stainless Steel
304 Stainless Steel
316 Stainless Steel
Epoxy - O-Ring Material:
- Model 50D: Viton
Model 50K: VitonÂ
Model 50S: NoneÂ
Model 50SD: None - Fitting Material:
- Model 50D: Choose from acetal, brass, or stainless steel
Model 50K: Choose from acetal, brass, or stainless steelÂ
Model 50S: Choose from acetal, brass, or stainless steelÂ
Model 50SD: Choose from acetal, brass, or stainless steel - Recommended Filtration:
- Model 50D: 20 microns or less
Optional inline filters available
Model 50K: 20 microns or less
Optional inline filters available
Model 50S: 20 microns or less
Optional inline filters available
Model 50SD: 20 microns or less
Optional inline filters available - Compatible Gases:
- Model 50D: Clean, dry gases compatible with wetted materials
Model 50K: Clean, dry gases compatible with wetted materialsÂ
Model 50S: Clean, dry gases compatible with wetted materialsÂ
Model 50SD: Clean, dry gases compatible with wetted materials - Output Signal:
- Model 50D: Minimum 2.5K ohm load
Not isolated
Model 50K: Minimum 2.5K ohm load
Not isolated
Model 50S: Minimum 2.5K ohm load
Not isolated
Model 50SD: Minimum 2.5K ohm load
Not isolated
Model 50SD: Minimum 2.5K ohm load
Not isolated - Warm-Up Time:
- Model 50D: Less than 5 minutes Model 50K: Less than 5 minutes Model 50S: Less than 5 minutes Model 50SD: Less than 5 minutes
- Integrated Display:
- Model 50D: 3.5 digit LCD flow rate display Model 50K: None Model 50S: None Model 50SD: 3.5 digit LCD flow rate display
- Typical Power Consumption:
- Model 50D: Standard: 12VDC @ 150 mA (12.5-15VDC)
E Suffix: 24VDC @ 80 mA (22-25VDC)
Model 50K: Standard: 12VDC @ 150 mA (12.5-15VDC)
E Suffix: 24VDC @ 80 mA (22-25VDC)
Model 50S: Standard: 12VDC @ 150 mA (12.5-15VDC)
E Suffix: 24VDC @ 80 mA (22-25VDC)
Model 50SD: Standard: 12VDC @ 150 mA (12.5-15VDC)
E Suffix: 24VDC @ 80 mA (22-25VDC) - Peak Power Consumption:
- Model 50D: Standard: 12VDC @ 300 mA (12.5-15VDC)
E Suffix: 24VDC @ 160 mA (22-25VDC)
Model 50K: Standard: 12VDC @ 300 mA (12.5-15VDC)
E Suffix: 24VDC @ 160 mA (22-25VDC)
Model 50S: Standard: 12VDC @ 300 mA (12.5-15VDC)
E Suffix: 24VDC @ 160 mA (22-25VDC)
Model 50SD: Standard: 12VDC @ 300 mA (12.5-15VDC)
E Suffix: 24VDC @ 160 mA (22-25VDC) - Electrical Connections:
- Model 50D: Integrated 36" (92 cm) cable, terminated with:
Standard: 6-pin Mini-DIN (PS/2 Style)
D1 Option: 9-pin D-Sub male
D2 Option: 15-pin D-Sub male
Model 50K: Integrated 36" (92 cm) cable, terminated with:
Standard: 6-pin Mini-DIN (PS/2 Style)
D1 Option: 9-pin D-Sub male
D2 Option: 15-pin D-Sub male
Model 50S: Integrated 36" (92 cm) cable, terminated with:
Standard: 6-pin Mini-DIN (PS/2 Style)
D1 Option: 9-pin D-Sub male
D2 Option: 15-pin D-Sub male
Model 50SD: Integrated 36" (92 cm) cable, terminated with:
Standard: 6-pin Mini-DIN (PS/2 Style)
D1 Option: 9-pin D-Sub male
D2 Option: 15-pin D-Sub male - Settling Time:
- Model 50D: Typically <1 second for 97% of final value
Model 50K: Typically <1 second for 97% of final valueÂ
Model 50S: Typically <1 second for 97% of final valueÂ
Model 50SD: Typically <1 second for 97% of final value - Reliability:
- Model 50D: 100,000 Hours MTBF (testing ongoing)
Model 50K: 100,000 Hours MTBF (testing ongoing)Â
Model 50S: 100,000 Hours MTBF (testing ongoing)Â
Model 50SD: 100,000 Hours MTBF (testing ongoing) - Approvals/Certifications:
- Model 50D: CE Approved
89/336/EEC (EN 55011 & EN 50082-1)
73/23/EEC Low Voltage Directive
Model 50K: CE Approved
89/336/EEC (EN 55011 & EN 50082-1)
73/23/EEC Low Voltage Directive
Model 50S: CE Approved
89/336/EEC (EN 55011 & EN 50082-1)
73/23/EEC Low Voltage Directive
Model 50SD: CE Approved
89/336/EEC (EN 55011 & EN 50082-1)
73/23/EEC Low Voltage Directive - Ratings:
- Model 50D: NEMA 1 (IP10)
Model 50K: NEMA 1 (IP10)Â
Model 50S: NEMA 1 (IP10)Â
Model 50SD: NEMA 1 (IP10) - Warranty:
- Model 50D: 1 year limited Model 50K: 1 year limited Model 50S: 1 year limited Model 50SD: 1 year limited
|
||||
|
||||
|
||||
|
||||
|
||||
|
||||
|
||||
|
||||
|
||||
|
||||
|